25 - Care and Maintenance
Cleaning the cuff sleeve, bladder and tubing
1. Read the safety instructions carefully and observe them closely before cleaning.
2. Before washing, carefully remove the bladder and tubing from the cuff sleeve.
3. When cleaning the cuff sleeve, bladder and tubing, use only mild detergents in lukewarm water without
fabric softener.
Attention
Damage to the cuff sleeve during washing
Always close the Velcro strip before washing!
It is possible to wash the cuff sleeve in the washing machine at max. 30°C. Do not spin.
Do not use fabric softeners or other washing aids (e.g. hygiene rinses, textile deodorants). These
agents can leave behind residues and damage the material.
The cuff sleeve is not suitable for drying in a dryer.

Disinfection

Caution
Intolerances caused by the use of disinfectants: Some patients display intolerance (e.g. allergies)
to disinfectants or their components.
Never use disinfectants which leave residues on the product or which are not suitable for contact with
the skin.
Carefully wash the cuff to remove residues.
Attention
Damage to the cuff sleeve, bladder and tubing caused by disinfectants
Do not submerge the cuff sleeve in disinfectants.
Avoid disinfecting the cuff bladder and connected rubber tubing.
The bladder and tubing can be damaged by disinfectants. Wipe down the bladder with lukewarm water
and add a mild detergent, if necessary.
Ensure that no liquid enters the tube opening.
The user (doctor) decides whether and when the ABP Monitor and the cuff sleeve should be disinfected for
hygienic reasons (e.g. after every use).
The following agents are recommended for disinfecting the ABP Monitor and the cuff sleeve:
Terralin Liquid (Manufacturer: Schülke & Mayr)
Isopropyl alcohol (70%)
For full effectiveness, moisten the ABPM 7100 and cuff sleeve with the disinfectant for at least 5 minutes.
The use of disinfectants not recommended in the directions for use shall render the user responsible for proof
of safe application.
Note
It is imperative that you observe the manufacturer's information regarding the use of these
products. Allow the agents to dry off completely.

Maintenance plan

Attention
Damage to device
Do not open the casing. Once the device is opened, all warranties will lapse.
Weekly Maintenance
Analysis review:
1. Review the print-out of your measurement analysis for:
Correctly entered times and intervals in accordance with the log.
Times of day/night transitions.
Correct standard values (nocturnal decrease).
2. Check the device, cuff and the cuff tubing for superficial soiling and clean it as specified in the Cleaning
section.
3. Check the cuff and the cuff tubing for superficial damage. In the event of damages return it to your
Welch Allyn specialist for inspection.
Checking battery voltage:
Always use fully charged or new batteries.
The battery voltage appears on the display of the ABPM 7100 for approximately 3 seconds after the device is
switched on. The battery voltage must be at least 2.6 volt to ensure a 24-hour measurement.
Maintenance every 2 years
As proof of continuous compliance to "Basic Requirements" pursuant to Directive 93/42/EEC, the ABPM 7100
must be subjected to calibration checks every two years. In certain countries, this requirement may be
regulated by national laws or regulations.
Welch Allyn offers to provide calibration checks and the servicing comprising of the following:
Calibration check
Software updates (if required)
Functional check: Electronics, pump and pneumatic circuit
Except the calibration check, no further maintenance work for electronic compatibility are necessary.

Troubleshooting

Attention
Damage to device
Do not open the casing. Once the device is opened, all warranties will lapse.

Basic error sources

The following may cause error measurements or unintended events:
The patient's arm movement during measurement
Incorrect cuff size
Cuff displacement while wearing it
Omitted successful initial measurement by the doctor
Wrong log set by the user
Empty, incorrectly charged or outdated batteries
Kinked or knotted cuff tubing
Severe arrhythmia