Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007, Table 6)
Recommended separation distances between portable and mobile RF
communications equipment and the electric breastpump Lactina
The electric breastpump Lactina is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the electric breastpump
Lactina can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the electric
breastpump Lactina as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output power of
transmitter W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separa-
tion distance d in metres (m) can be estimated using the equation applicable to the frequency
of the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affec-
ted by absorption and reflection from structures, objects and people.
Separation distance according to frequency of
transmitter M
150 kHz to 80 MHz
d = 1,167√P
0,12
0,37
1,17
3,69
11,67
80 MHz to 800
MHz
d = 0,35√P
0,04
0,07
0,11
0,22
0,35
0,7
1,11
2,21
3,50
7,00
800 MHz to 2.5 GHz
d = 0,7√P
169