12. EMC / Technical description
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
U
NOTE
is the a.c. mains voltage prior to application of the test level.
T
166
<5 % U
T
(>95 % dip in U
)
T
for 0,5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(>95 % dip in U
T
for 5 s
3 A /m
<5 % U
T
(>95 % dip in U
)
T
for 0,5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(>95 % dip in U
)
T
for 5 s
30 A /m
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the
electric breastpump
Lactina requires
continued operation
during power mains
interruptions, it is
recommended that
the electric breast-
pump Lactina is
powered from an
uninterruptible power
supply or a battery.
Power frequency
magnetic fields should
be at levels charac-
teristic of a typical
location in a typical
commercial or hospi-
tal environment.