Option
Primary Interval
Secondary Interval Enables a second time interval between measurements to become active when the
Secondary Interval
Condition
User
Location Name
Office
7. To start the log, push the RIGHT arrow. The log becomes active. The amount of time before the
first measurement is done shows.
Copy the logs to the instrument
Logs must be saved from the sonde to the instrument before they can be shown on the instrument
display.
Note: Only completed log files can be saved to the instrument.
1. Push
until three or four dots show at the top of the screen.
2. Push the RIGHT arrow until "Sonde HL" shows at the top of the display.
3. Select Logging. The status and details of the last log made show.
4. To copy the last log made to the instrument, push OK.
5. To copy a different log to the instrument, push
6. Select Log Files on Sonde HL.
7. Select the log file to copy.
8. Select Transfer Log File.
Look at a log file
Log files show on the display in the same graph format as real-time measurements and depth profile
measurements.
Note: Logs must be saved from the sonde to the instrument before they can be shown on the instrument display.
Refer to
Copy the logs to the instrument
1. Push
until three or four dots show at the top of the screen.
2. Push the RIGHT arrow until "Surveyor HL" shows at the top of the display.
3. Select Log Files.
4. Select a log file, then select Show Details. The details of the log file show.
5. Push OK to show the graph.
Description
Sets the time interval between measurements (1 second minimum). Make sure that the
primary interval is longer than the sensor warm-up time.
Note: For deployment with the mooring cap, the time interval affects the battery life.
The longer the time interval the longer the battery life. The life of a new battery is
approximately 75 days of use with a 15-minute logging interval and a 30-second warm-
up time with temperature, conductivity, pH and LDO sensors installed.
selected parameter(s) (e.g., temperature) is within the selected range. A maximum of
four triggers can be selected at one time. The primary interval becomes inactive when
the secondary interval is active.
The secondary interval is typically shorter than the primary interval. Make sure that the
secondary interval is longer than the sensor warm-up time.
Sets the parameter range(s) that makes the secondary interval active.
Adds a user name to the log.
Adds a location name to the log.
Adds an office name to the log.
on page 21.
.