5. To change the view options, push
Note: The view options settings do not change what is recorded to log files. All measurements are recorded to
log files.
Option
Description
Parameters
Sets the parameters that show on the display. Push OK to select or not select a
parameter.
Parameter
Sets the sequence the parameters show in the list view.
Order
To move a parameter up or down the list, select the parameter, push OK and then push
the UP or DOWN arrow.
Graph Order
Sets the parameters that show in the graph view. Only the first two parameters in the list
show in graph view.
To move a parameter up or down the list, select the parameter, push OK and then push
the UP or DOWN arrow.
Grid Order
Sets the parameters that show in the grid view and the locations of the parameters in the
grid. Only the first four parameters in the list show in grid view.
To move a parameter up or down the list, select the parameter, push OK and then push
the UP or DOWN arrow.
Update Interval
Sets the time interval between measurements.
Stability Check
Sets the stability check to on or off. When set to on, a status icon shows at the bottom of
the display that identifies if the measurement is stable or not stable. Refer to
on page 10 for status icon descriptions. Measurements that are not stable are identified in
the log files.
Note: A measurement is identified as stable when it is within the selected stability criteria
for the parameter.
Stability
Sets the stability check criteria used by the software to identify when measurements are
Criteria
stable.
Select a parameter and push OK. Push
Max Delta—The maximum difference between the current measurement and the average
measurement for a measurement to be identified as stable. Number of samples—the
number of measurements that are used to calculate the average measurement used for
stability checks.
Note: Stability check settings are not used for calibrations.
6. To go out of real-time monitoring, push
Figure 8 Real-time monitoring views
Save real-time measurements
Real-time measurements are manually saved to a log file(s) on the instrument.
. Select an option, then push OK.
to set:
.
Status icons
English 17