en
Precaution
The electrodes are intended for single
use only. Safely discard after use to
prevent injury to hospital personnel. These
electrodes are not designed to withstand
resterilization. Do not resterilize.
Notice
This electrode has a coating to reduce
sticking of eschar. Cleaning the electrode
with a scratch pad or other abrasive object,
scraping with a sharp object, or bending
beyond 90 degrees may damage the
electrode. If the electrode is damaged,
discard it.
Using coated electrodes at high power
settings may cause damage to the coating.
If the coating is damaged, discard the
electrode.
Electrosurgical generators (e.g., Force FX™
or Force EZ™ generators) produce desired
surgical effects at lower power cut mode
settings than conventional electrosurgical
generators. The electrode coating may
deteriorate when used with tissue response
generators at higher power settings.
6
Important
Wipe the electrode often with moist gauze
or other material.
Instructions for Use
Ensure the pencil is not connected to the
1.
generator.