Warning
• Firmly grasp as much of the hemostat as
possible before activating the generator.
This disperses the current over a larger
area and minimizes the current
concentration at the fingertips.
• "Buzz the hemostat" below hand level
(as close as possible to the patient) to
reduce the opportunity for current
to follow alternate paths through the
surgeon's hands.
• When using a coated or nonstick blade
electrode, place the edge of the electrode
against the hemostat or other metal
instrument.
Precaution
Always use the lowest power setting that
achieves the desired surgical effect. Use
the active electrode for the minimum time
necessary in order to reduce the possibility
of unintended burn injury.
Precaution
Pediatric applications and/or procedures
performed on small anatomic structures
may require reduced power settings. The
higher the current flow and the longer the
current is applied, the greater the possibility
of unintended thermal damage to tissue,
especially during use on small appendages.
Before use, examine the electrosurgical unit
and accessories for defects. Do not use
cables or accessories with damaged
(cracked, burned, or taped) insulation
or connectors.
Do not modify or add to the insulation of
active electrodes.
Activate the electrosurgical unit only when
ready to deliver electrosurgical current and
the active tip is in view (especially if looking
through an endoscope).
Deactivate the electrosurgical unit before
the tip leaves the surgical site.