Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

6.4 Troubleshooting

Important!
The following table shows a number of common malfunctions and their probable
cause. If a malfunction cannot be eliminated by following the instructions in the
table, please contact your authorised dealer.
Problem
No motor noise/lift does not
move
Lift can be moved upwards but
not lowered
Lift stops during lifting opera-
tion
Battery cannot be recharged
The bathlift is sliding in the
bathtub
19
Possible causes
Battery is not properly inserted
Battery is empty
Contacts are contaminated
Contacts are bent
Cable is damaged
Low-voltage protection is trig-
gered
Load is excessive
Fuse is defective
Battery is defective
Suction cups are worn or dam-
aged
Remedy
Insert battery (see chapter 5.3)
Recharge battery (see
chapter 5.1)
Clean contacts
Readjust contacts
Contact vendor for replace-
ment of the cable
Recharge battery
The AQUATEC FORTUNA
may only be used by persons
weighing less than 120 kg.
Contact your dealer to check
the battery
Replace battery
Replace suction cups
EN

Publicité

Chapitres

Table des Matières
loading

Table des Matières