Boston Scientific WaveWriter Alpha Mode D'emploi page 15

Masquer les pouces Voir aussi pour WaveWriter Alpha:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 20
1. Fully insert the Leads, Lead Extensions, Splitters, and/or Connector(s) into the IPG Ports, being careful
not to stress or bend the proximal end of the Lead. When the Lead is properly inserted, the Lead will
stop and the retention ring will be located under the Set screw.
Note: Lead Extensions, Adapters, and Splitters are not full-body MRI scan eligible.
2. Fully insert a Port Plug into unused IPG Ports.
Note: If you experience difficulty when inserting the Lead, Lead Extension, Splitter, Connector, or Port
Plug, use the Hex Wrench to loosen (counterclockwise) the Setscrew and/or gently rotate the Lead
to help advance the proximal end.
Note: To confirm good connections, check impedances before tightening the Setscrew. The IPG must be
in contact with the subcutaneous pocket in order to receive accurate impedance measurements.
3. Pass the Hex Wrench through the hole in the septum located on the front or back of the IPG header
and tighten each Setscrew until the Hex Wrench "clicks," indicating lock.
CAUTION: Ensure that the Lead is fully inserted before tightening the Setscrew to prevent Lead
damage
Note: If a Port Plug is used, it is still necessary to tighten the Setscrew on the Port Plug as described
above.
Note: The hex wrench is torque-limited and cannot be over-tightened.
4. Place the IPG in the subcutaneous pocket with logo marking facing up towards the skin.
5. Coil excess Lead, Extension, Splitter, or Connector under the IPG.
6. If desired, secure the IPG in the pocket by suturing through the holes in the IPG header.
CAUTION: Do not Suture through the Leads or Splitter.
CAUTION: Tightening Sutures directly on the Lead can damage the Lead. In addition, during an MRI
scan, the patient could be at risk of tissue heating, resulting in tissue damage or serious patient injury.
7. Close and dress the wound(s).
Instructions for Use
Implantable Pulse Generator
92469342-02
11 of 353

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Wavewriter alpha prime

Table des Matières