Calibration And Adjustment - Thermo Fisher Scientific Thermo Scientific Finnpipette F1 Guide D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Calibration and adjustment

All Finnpipettes are factory calibrated and adjusted to give the volumes as specified with
distilled or deionized water using the forward pipetting technique. It should be noted that the
use of other pipetting techniques may affect the calibration results. The pipettes are constructed
to permit re-adjustment for other pipetting techniques or liquids of different temperature and
viscosity.
Device requirements and test conditions
An analytical balance must be used. The scale graduation value of the balance should be chosen
according to the selected test volume of the pipette:
Volume range
under 10 µl
10-100 µl
above 100 µl
Test liquid: Water, distilled or deionized, "grade 3" water conforming ISO 3696. Tests are done
in a draft-free room at a constant (±0.5°C) temperature of water, pipette and air between 15°C
to 30°C. The relative humidity must be above 50%. Especially with volumes under 50 µl the
air humidity should be as high as possible to reduce the effect of evaporation loss. Special
accessories, such as the evaporation trap, are recommended.
Procedure to check calibration
The pipette is checked with the maximum volume (nominal volume) and with the minimum volume.
A new tip is first pre-wetted 3-5 times and a series of ten pipettings is done with both volumes.
A pipette is always adjusted for delivery (Ex) of the selected volume. Use of forward pipetting
technique is recommended. The maximum permissible errors are designed for forward method.
Procedure:
1. Do 10 pipettings with the minimum volume.
2. Do 10 pipettings with the maximum volume.
3. Calculate the inaccuracy (A) and imprecision (cv) of both series.
4. Compare the results to the limits in the Table 1.
If the calculated results are within the selected limits, the adjustment of the pipette is correct.
TABLE 1: Maximum permissible errors according ISO8655
Range
Volume
0,2-2 µl
0,5-5 µl
1-10 µl
2-20 µl
5-50 µl
10-100 µl
20-200 µl
30-300 µl
100-1000 µl
0,5-5 ml
1-10 ml
readable graduation
0.00 1 mg
0.01 mg
0.1 mg
Inaccuracy
µl
µl
2
±0.080
0.2
±0.080
5
±0.125
0.5
±0.125
10
±0.120
1
±0.120
20
±0.20
2
±0.20
50
±0.50
5
±0.50
100
±0.80
10
±0.80
200
±1.60
20
±1.60
300
±4.0
30
±4.0
1000
±8.0
100
±8.0
5000
±40.0
500
±40.0
10000
±60.0
1000
±60.0
%
s.d. µl
±4
0.040
±40
0.040
±2.5
0.075
±25
0.075
±1.2
0.080
±12
0.080
±1.0
0.10
±10.0
0.10
±1.0
0.20
±10
0.20
±0.8
0.30
±8.0
0.30
±0.8
0.60
±8.0
0.60
±1.3
1.5
±13
1.5
±0.8
3.0
±8.0
3.0
±0.8
15.0
±8.0
15.0
±0.6
30.0
±6.0
30.0
7
Imprecision
cv%
2.0
20.0
1.5
15
0.8
8.0
0.5
5.0
0.4
4.0
0.3
3.0
0.3
3.0
0.5
5.0
0.3
3.0
0.3
3.0
0.3
3.0

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Finnpipette f1

Table des Matières