Calling Up The Menu Service Functions; Error Messages On The Display - Aesculap Nelson delux GN 640 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 64
High frequency surgical device
On Phase memory
The On Phase memory saves the last 100 on phases of the HF device.
The following parameters are shown:
• Date (DD.MM.YY)
• Switch-on time
• Duration of on phase (hh:mm)
• Type of neutral electrode (1 = type 1, 2 = type 2, 3 = type 3)
• Maximum HF switching time
• Instrument recognition setting (0 = off, 1 = default,
2 = last settings)
• autoCOAG automatic switch-off setting (S = short, L = long)
• Argon allocation (0 = none, 1 = monopolar 1, 2 = monopolar 2)
• Highest occured temperature of the monopolar power amplifier

7.2 Calling up the menu Service Functions

The following data can be retrieved in the Service Functions menu:
• Circuit board identification:
shows art. no., serial no. and the status of any alterations made in
the integrated circuit board
• Actual value:
shows various analog and digital actual values in the microcontrol-
ler periphery (see Service Manual)
• Service mode:
OFF: Error list (see Error messages on the display)
ON: Error list (see Service Manual)
Press function key
2.
The display shows:
Note
After a waiting time of 30 seconds, or when the user presses function
key
, the system returns to the main menu.
(GN 640)
Press function key
3 to call up the menu Service Functions.
Press function key
to call up the required service function or
to activate (ON) or deactivate (OFF) the service mode.
Confirm the setting by pressing function key
main menu.
- or -
To exit the menu without adopting the changes:
Press the function key

7.3 Error messages on the display

If the device detects errors in the course of a self-test, these errors are
displayed with an error number, a pictogram and an error text.
Depending on the error detected, the device disables the monopolar or
bipolar HF outputs or the monopolar and bipolar HF outputs.
and return to the