7. Recognizing and remedying errors
7.1 Calling up the activation, error and On Phase
memories
Press function key
2.
The display shows:
Note
After a waiting time of 30 seconds, or when the user presses function
key
, the system returns to the main menu.
Press function key
4 to call up the menu History Record.
A pop-up menu is displayed.
Press function key
to select the required memory menu.
Confirm the setting by pressing function key
- or -
Leave the menu:
Press the function key
Keys
are used for going from page to page within the
menu; keys
are pressed to jump to the beginning or the
end of the list.
Activation memory
The activation memory saves the last 600 activations of the HF device.
The following parameters are shown:
• Date of activation (DD.MM.YY)
• Time of activation (hh:min)
• Duration of activation (in seconds)
• Output
– BI1: bipolar 1
– BI2: bipolar 2
– MO1: monopolar 1
– MO2: monopolar 2
• Dosage
54
High frequency surgical device
.
or wait for 30 seconds.
• Mode
– CUT0: POWERCUT 1/BICUT 1 COAG0: CONTACTCOAG / BICOAG
– CUT1: POWERCUT 2/BICUT 2 COAG1: SOFTCOAG
– CUT2: POWERCUT 3
– CUT3: POWERCUT 4
– CUT4: MICROCUT 1
– CUT5: MICROCUT 2
– CUT6: SEALCUT
• Activation mode on/off (H: hand, A: automatic, F: foot pedal)
2
2
• I
*t-value (in 0.1 A
s)
• Instrument used
– unknown: Instrument without instrument recognition
– seven-digit number: 3rd to the 5th digit of the article number,
and the last 4 digits of the serial number of the instrument
Error memory (Error by date; Error by number)
The error memory saves the last 100 error messages generated by the
HF device. High-priority errors are saved for a longer period. The fol-
lowing parameters are shown:
• Date (DD.MM.YY)
• Time
• Error number
• Error text
(GN 640)
COAG2: SPRAYCOAG
COAG3: FORCEDCOAG