High frequency surgical device
Selecting a dosage
A dosage must be set for each HF output that is selected.
The allowable dosage range depends upon the operational mode se-
lected:
Operational mode
Monopolar
MICROCUT 1
MICROCUT 2
POWERCUT 1
POWERCUT 2
POWERCUT 3
POWERCUT 4
SEALCUT
SOFTCOAG
CONTACTCOAG
FORCEDCOAG
SPRAYCOAG
Bipolar
BICUT 1
BICUT 2
BICOAG
The size of the graduations between dosages depends upon operation-
al mode and dosage range:
Operational mode
Dosage range
Monopolar CUT
1—50
55—100
110—300
Monopolar COAG
1—50
55—120
Bipolar CUT
1—50
55—100
Bipolar COAG
0,1—10
10,5—20
21—50
55—100
To select HF output for which the dosage is to be selected: Press
the corresponding function key
Dosage min.
Dosage max.
5
300
5
300
5
300
5
250
5
200
5
150
5
100
5
100
1
120
1
120
1
120
1
100
1
100
0,1
Foot pedal: 100
autoCOAG: 50
Graduation
1
5
10
1
5
1
5
0,1
0,5
1
5
3, 4, 5, 6.
(GN 640)
Set the dosage for each HF output, using function keys
and
.
Note
An additional bar icon appears in the display showing where the select-
ed dosage lies within the allowable range.
Selecting activation mode foot pedal/hand switch/auto-
COAG
An activation mode must be set for each output that is activated: with
the foot pedal, the hand switch or automatically.
Icons on menu bar activation mode:
Description
HF activation and switch-off via the
foot pedal
HF activation via the foot pedal and
automatic HF switch-off
HF activation and switch-off via the
hand switch
HF activation via the hand switch and
automatic HF switch-off
Automatic switch-on/off for coagula-
tion (only available in bipolar mode);
Note
- only available in bipolar mode
- the maximum dosage setting is
reduced to 50
45