Interconnections

The following figure shows the TwisTorr 305 FS interconnections.
Figure 25

Earth (Ground) Connection

We recommend that you fit a separate earth (ground) conductor to the
TwisTorr 305 FS. Use a separate insulated green/yellow conductor, and use
a M5 x 10 screw and shake proof washer (fitted to the earth hole on the pump) to
secure the earth conductor to the pump. The screw must be fixed with a 2 Nm torque.
The impedance between the pump-body and the earth connection point must
be < 0.1 Ω.
TwisTorr 305 FS User Manual / 87-901-057-01 (A.00)
Technical Information
EARTH GROUND
CONTROLLER
CONNECTOR

Interconnections

FORELINE
233/258