Technical Information
Technical Specification
CHARACTERISTIC
Cooling water
Bakeout temperature
Vibration level
(when water cooled)
(Data measured with an
indicated bearing temperature of
20 °C: see vibration graph in the
following figure)
Lubricant
Noise level
Power supply:
Protection fuse
Compliance with:
Installation category
Pollution degree
Power cable
214/276
Input voltage:
Input freq.:
Max input power:
Stand-by power:
Max operating
power:
TV 551/701 SEM Navigator User Manual /
TV 551 SEM
flow: 200 l/h (0.89 GPM)
temperature: + 10 ° C to + 30 °C
pressure: 3 to 5 bar (45 to 75 Psi)
120 °C at inlet flange max. (CF flange)
80 °C at inlet flange max. (ISO flange)
bearing peak: < 45 mm/sec
residual unbalance: < 350 mm/sec
permanent lubrication
<45 dB(A) at 1 meter
100 to 240 Vac
50/60 Hz
500 VA
30 to 35 W
TV 551 SEM:
325 W (water cooling)
220 W (air cooling)
TV 701 SEM:
350 W (water cooling)
250 W (air cooling)
1 x 6.3 A
UNI EN 292-1
UNI EN 292-2
EN-CENELEC 55011
IEC 1000-4-2 (ex 801-2)
IEC 1000-4-3 (ex 801-3)
IEC 1000-4-4 (ex 801-4)
EN 61010-1 (IEC 1010-1)
EN 1012-2
II
2
With European or NEMA plug
3 meters long (optional)
TV 701 SEM
2
2
87-900-958-01(G)