Guidance and manufacturer's declaration – electromagnetic emissions
BRIZZY is intended for use in the electromagnetic environment specified below. The customer
or the user of BRIZZY should assure that it is used in such an environment.
Emissions test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic
emissions
IEC 61000-3-2
Voltage fluctu-
ations/ flicker
emissions
IEC 61000-3-3
© Nomics s.a. All rights reserved
Oximeter (Nonin Xpod)
Measurement range: 0 to 100%
Accuracy (for adults using the 8000AA Finger Clip Sensor, in 70 to
100% range, with ±1SD): ±2%
Sampling frequency: 3Hz
Conditioning: 4-beat Average
Measurement range: 18 to 300bpm
Accuracy (for adults using the 8000AA Finger Clip Sensor,
no motion): ±3bpm
Sampling frequency: 3Hz
Conditioning: 4-beat Average
Measurement Range: ±5.3mV
Accuracy: 2.6 μV
Sampling frequency: 125Hz
Conditioning: 0,15Hz r 40Hz
Measurement Range: 0 to 200bpm
Accuracy: ±5% up to 150bpm, ±7.5% worst case
Sampling frequency: 3Hz
Conditioning: 3-beat Average
Compliance
Electromagnetic environment – guidance
BRIZZY uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
Group 1
likely to cause any interference in nearby electronic
equipment.
Class B
BRIZZY is suitable for use in all establishments other
Not applicable
than domestic and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
Not applicable
ECG module
MU_BRIZZY_REV5 / 03-2017 Eng-33