Problem
The yellow
indicator light is on
while the device
is connected to a
computer.
The computer
does not detect the
device although
the device is
connected to the
computer.
The start date/time
of the recording
does not match
what is expected.
© Nomics s.a. All rights reserved
Possible cause
The yellow indicator light being
on is normal in these conditions.
It indicates that the device has
detected the presence of a USB
connection.
If the yellow indicator light on the
front of the device does not light
up, it is likely that a plug is not
correctly inserted into its socket.
The computer has not detected
the device correctly.
Last recording did not start and
the internal memory still contains
previous recordings.
Time of the recorder is not
correct.
Solution
-
Reinsert the plug of the special
USB cable in the white socket
of the device. Reinsert the USB
plug of the cable into a USB port
on the computer.
If the device does not appear
in the import utility list, use the
'Refresh' button on the right. If
necessary, unplug the device
from the computer and replug
it. If it still does not work, restart
the computer and try again.
Restart the recording.
Perform a time setting by simply
connecting the device to the
computer and launching CERES.
The recordings that follow
should indicate the exact date.