Télécharger Imprimer la page

Mounting The Pipette Clamp; Checking The Leak Tightness; Replacing The Rechargeable Battery - StarLab ErgoOne Fast Mode D'emploi

Masquer les pouces Voir aussi pour ErgoOne Fast:

Publicité

Les langues disponibles

Les langues disponibles

6.3 Mounting the pipette clamp

6.4 Checking the leak tightness

1. Insert the pipette.
2. Fill the pipette with water.
3. Hold the pipette vertically.
4. Observe the pipette outlet for approx. 30 seconds.
Do not touch the pipette. Do not press the control buttons.
No water may be allowed to escape.
5. If water escapes, disassemble and carefully reassemble the pipette clamp.

6.5 Replacing the rechargeable battery

Proceed as follows:
1. Slide the battery compartment lid open.
2. Remove the battery.
3. Insert a new battery.
1.
Push the seal with the groove up into the
filter adapter.
2.
Push the wide opening of the membrane
filter into the narrow opening of the
pipette adapter.
3.
Guide the aspirating cone over the pipette
adapter and turn it until it engages.
ErgoOne FAST · User Manual ·
27

Publicité

loading