6
MAINTENANCE
6.1 Disassembling the pipette clamp
If liquid has entered the pipette clamp, the aspiration capacity may be decreased, or pipette
clamp assemblies may be damaged. The pipette clamp must be disassembled in order to clean
or replace the assemblies.
1. Turn the aspirating cone counterclockwise and remove it.
2. Remove the pipette adapter and membrane filter from the filter adapter.
3. Remove the membrane filter from the pipette adapter.
4. Use a narrow pointed object to pry the seal out of the filter adapter.
6.2 Cleaning
6.2.1
Cleaning the Pipette Controller
Special service is not required.
NOTICE! Damage to the device due to autoclaving.
›
Do not autoclave the pipette controller.
To clean contaminated surfaces, proceed as follows:
›
Wipe the housing using a damp cloth.
›
Disinfect surfaces using alcohol (ethanol, propanol) or alcohol-containing disinfectants.
ErgoOne FAST · User Manual ·
25