Operation
•
Calibration of buffer 2 starts 2 sec after pressing the keys CAL, "UP" or
"DOWN", as above; marker "P2" is displayed steadily.
•
After successful calibration the LCD elements "P1" and "P2" are switched
off and the new calibrating data is displayed:
•
•
•
The meter display returns to the measuring operation previous to the
calibration. If the temperature was being measured, the parameter pH is
selected.
6.4.2
•
Press the key "CAL" followed by the key "pH/mV/°C" to display the
calibrating data again later.
•
The meter display returns to the measuring operation previous to the
calibration. If the temperature was being measured, the parameter pH is
selected.
6.4.3
•
Press the key "pH/mV/°C" after successfully calibrating buffer 1 to
discontinue the calibration process.
•
The calibration is now interpreted as a 1-point calibration and new
calibrating data are displayed. Depending on the value of the selected
buffer, the zero point or slope are calibrated:
Selected buffer value:
New calculation of:
36
zero point for 2 sec with activated LCD elements "cal" and "∆pH", and
afterwards
slope for 2 sec with activated LCD elements "cal" and "%".
Displaying calibration data
One point calibration
selected buffer value
new calculation of:
0.00 – 5.49
slope
5.50 – 8.50
8.51 – 14.00
zero point
slope