ProMinent DULCOMETER pH/mV/Pt 1000 Mode D'emploi page 33

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

5.2.2
Operating the key CAL always starts a pH calibration independent of the
current operating state. If no calibration is being carried out at the time, a new
calibration is started, otherwise the current calibration is started afresh.
5.2.3
The simulation mode is only activated if the temperature probe is not
connected to the right-hand SN-6 connector.
before
SIM pH
SIM Pt 1000
SIM redox
Measure pH
Measure Pt 1000:
cannot be carried
out without
Pt 1000 probe
CAL pH
®
ProMinent
Function key "CAL"
Function key "SIM"
Operating state
subsequent
Manual
temperature
compensation
SIM redox
SIM pH
SIM pH
------
CAL pH
Function Description
Description
Switches from pH simulation to adjusting
the manual temperature correction value
(conforms to the manual temperature
compensation) for the pH measurement.
Switches over from manual temperature
compensation to redox simulation.
Switches over from redox simulation to
pH simulation.
Switches from operating state measure
pH to SIM pH.
Simulation not selectable with Pt 1000
probe connected.
Remains in calibration state.
33

Publicité

Table des Matières
loading

Table des Matières