Press TARE and the display will change to numeric data input.
The temperature currently set will be displayed if water, ethanol or methanol is
selected.
On selecting "OTHER" the currently set density of the fluid is displayed.
either
during selection of water, ethylic or methyl alcohol read temperature at the
thermometer and enter with the help of the navigation keys.
or
if selecting "OTHER" enter density of the auxiliary liquid of your choice.
Numeric input
# indicates that the weighing balance is in numeric input mode. The first digit is
flashing and can be changed.
Increase flashing digit
Decrease flashing digit
Digit selection to the right
Confirm entry
The weighing balance will display the auxiliary fluid's density at the entered
temperature for about 3 sec and will then change to "density determination solid
matter" mode.
To change from density mode to weighing mode press MENU for 3 s.