KERN YDB-03 Mode D'emploi page 117

Kit universel de détermination de la densité pour balances analytiques
Masquer les pouces Voir aussi pour YDB-03:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

 Place sample of solid material in the upper sample dish, see fig. 1, chap. 5.1.2.
 Wait for stability display [ ], then take over the weight value using the PRINT
button.
 Wait until the display for determination of „sample in aid liquid" appears. Remove
the sample and if required tare by using the TARE button.
 Lay the sample into the lower sample dish and immerse it in the aid liquid trying to
avoid bubble formation.
Make sure that the sample is at least 1 cm immersed.
 Wait for stability display [ ], take over the weight value using PRINT button. The
density of the sample is shown.
 When an optional printer is connected, the displayed value can be edited using
the PRINT button.
Printout example (KERN YKB-01N):
To avoid corrosion, don't leave the density set immersed in liquid for a long time.
YDB-03-BA-e-1410
07-01-14
09:35:17
3
d: 8.0700 g/cm
39