5.2.2 Enter density of the auxiliary liquid
In weighing mode press the CAL key repeatedly until „SettinG" is displayed.
Press the TARE key.
Press the CAL-key repeatedly until „LSG Set" is displayed.
Press the TARE key, the currently set density of the auxiliary liquid is displayed.
In the upper part of the display panel, the
in order to indicate numerical input status. The active digit is flashing.
To change with the navigation keys enter density of your auxiliary liquid,
see chap. 8.
UNIT key:
Increase the flashing digit
PRINT] key:
Digit selection to the right
TARE key:
Confirm entry
Back to menu / weighing mode, press ON/OFF key repeatedly