 Wait until the display for weight determination of „sinker in test liquid" appears.
 If possible, immerse the sinker bubble-free in the test liquid.
Make sure that the sinker is at least 1 cm immersed (see fig. 6, chap. 6.1.2).
 Wait for stability display [ ], take over the weight value using PRINT button. The
density of the test liquid is displayed.
 When an optional printer is connected, the displayed value can be edited using
the PRINT button.
Printout example (KERN YKB-01N):
If at the density determination errors have appeared, „d-----„ is displayed.
 For further measurements go back to density determination mode, press MENU
button.
 Back to weighing mode, press ON/OFF button.
To avoid corrosion, don't leave the density set immersed in liquid for a long time.
YDB-03-BA-e-1410
07-01-14
09:35:17
3
d: 0.9984 g/cm