Method 2:
Make sure that the balance is in the mode for density determination of solid
material (see chap. 5.2.1).
If weighing balance does not show Zero, press TARE.
Place solid material into the upper sample dish see fig. 3, chap. 5.1.3.
The weight of the sample in air is displayed
Wait for stability display ( ), then press the CAL key.
Place the solid material entirely under the lower sifting bowl, see fig. 3, chap.
5.1.3.
For this purpose remove the immersion basket and for re-immerging keep the
sample possibly free of bubbles. Or if possible place the sample using pincers or
alike directly under the sifting bowl.
The weighing balance will first determine then display the sample's density.
If you connect an optional printer you can print the result.
Remove the sample. To carry out further measuring, press the CAL-key and start
with step 2.
To avoid corrosion, don't leave the density set immersed in liquid for a long time.