Method 2:
 Remove the immersion basket and place the glass beaker filled with the auxiliary
liquid in the centre of the platform. Filling height should be approx. ¾ of the
capacity. Make sure that it has no contact with the frame.
Remount the immersion basket. Make sure that it does not touch the glass
beaker. Set balance to zero.
 Make sure that the balance is in the mode "density determination solid material"
(see chap. 5.1.1).
 Put the solid material into the upper sample dish.
The weight of the „sample in air" is displayed.
 Wait until stability display ( ) appears, then press UNIT. "SINK" will be shown.
YDB-03-BA-e-1410
Fig. 3: Weighing in air
23