Method 2:
Remove the immersion basket and place the glass beaker filled with the auxiliary
liquid in the centre of the platform. Filling height should be approx. ¾ of the
capacity. Make sure that it has no contact with the frame.
Remount the immersion basket. Make sure that it does not touch the glass
beaker. Set balance to zero.
Make sure that the balance is in the mode "density determination solid material"
(see chap. 5.1.1).
Put the solid material into the upper sample dish.
The weight of the „sample in air" is displayed.
Wait until stability display ( ) appears, then press UNIT. "SINK" will be shown.
YDB-03-BA-e-1410
Fig. 3: Weighing in air
23