5.4 Series KERN ALS-A
5.4.1 Determine density of sinking solid bodies (d >1 g/cm
Remove the immersion basket and place the glass beaker filled with the auxiliary
liquid in the centre of the platform. Filling height should be approx. ¾ of the
capacity. Make sure that it has no contact with the frame.
Remount the immersion basket. Make sure that it does not touch the glass
beaker.
In weighing mode press MENU button. The first menu item „count" is displayed.
Press MENU button
Acknowledge using PRINT button, the current setting is displayed.
Using MENU button select „d SoLid"
Confirm by pressing the PRINT button. The currently set density of the aid liquid
is displayed (factory setting 1.0000 g /cm
To change, enter the density of the aid liquid using arrow keys .
Confirm input by pressing the PRINT button.
The display for weight determination of the "sample in air" appears.
Confirm by pressing the PRINT button.
Should the balance not show Zero, press the TARE button.
Place sample of solid material in the upper sample dish, see fig. 1, chap. 5.1.2.
Wait for stability display ( ), then press
Wait until the display for weight determination of „sample in aid liquid" appears.
Confirm by pressing the PRINT button.
YDB-03-BA-e-1410
3
)
3
for distilled water at 20°C).
.
35