6.3.2 Determining density of the test liquid

 Fill test liquid into the glass beaker.
Make sure that the balance is in the mode for density determination of liquids (see
chap. 6.3.1).
 Place sinker into the upper sample dish, see fig. 5, chap. 6.1.2.
The weight of the sinker in air will be displayed.
 Wait for stability display (
 Place sinker into the lower sifting bowl, see fig.6, chap. 6.1.2..
For this remove the immersion basket out from the frame. Always ensure that,
when re-immersing into the liquid, no additional bubbles adhere; it is better to use
pincers or alike to place the sample directly on the sifting bowl.
 Wait for stability display (
The weighing balance will first determine then display the fluid's density.
 When connecting an optional printer, the result will be printed out.
Printout example KERN YKB-01N:
To avoid corrosion, don't leave the density set immersed in liquid for a long time.
YDB-03-BA-e-1410
If necessary, set balance to zero.