Please clean the plug of power cord at least once a year.
Too much dust on plug may cause the fire.
The following safety checks should be performed at least every 12 months by a
qualified person who has adequate training, knowledge, and practical experience
to perform these tests.
The data should be recorded in an equipment log. If the device is not functioning
properly or fails any of the following tests, the device has to be repaired.
1) Inspect the equipment and accessories for mechanical and functional damage.
2) Inspect the safety relevant labels for legibility.
3) Verify that the device functions properly as described in the instructions for
use.
4) Test
the
earth
500μA, SFC: 1000μA.
5) Test the enclosure leakage current according to IEC 60601-1: Limit: NC
100μA, SFC: 500μA.
6) Test the patient leakage current (normal operation) according IEC 60601-1:
Limit: type CF: for a.c.: 10μA, for d.c.: 10μA.
7) Test the patient leakage current under single fault condition according IEC
60601-1: Limit: type CF: for a.c.: 50μA, for d.c.: 50μA.
8) Test the patient leakage current Mains voltage on applied part: According IEC
60601-1: Limit: type CF: for a.c.: 50uA.
Warning: No use-serviceable parts inside, before
servicing to authorized representative or manufacturer.
leakage
current
98
www.spengler.fr
according
IEC 60601-1: Limit: NC