Guidance and manufacture's declaration – electromagnetic immunity – for all
Guidance and manufacture's declaration – electromagnetic immunity
The MASTER PALM 3® Handheld pulse oximeter is intended for use in the
electromagnetic environment specified below. The customer or the user of
MASTER PALM 3® Handheld pulse oximeter should assure that it is used in
such an environment.
IEC 60601
Immunity test
test level
Electrostatic
discharge
± 6 kV
(ESD)
contact
IEC 61000-4-2
± 8 kV air
Electrical
± 2 kV for
fast transient/
power supply
burst
lines
IEC 61000-4-4
± 1 kV for
signal line
± 1 kV
Surge
differential
mode
IEC 61000-4-5
± 2 kV com-
mon mode
Voltage
dips, short
interruptions
< 5 % U
and voltage
(> 95 % dip
variations on
in U
power supply
cycle
input lines
IEC 61000-4-11
EQUIPMENT and SYSTEMS
Compliance
level
± 6 kV contact
± 8 kV air
± 0,5 kV kV
for power
supply lines
± 0,5 kVfor
signal line
± 1 kV
differential
mode
± 2 kV com-
mon mode
< 5 % U
T
(> 95 % dip
) for 0,5
in U
) for 0,5
T
T
cycle
108
www.spengler.fr
Electromagnetic envi-
ronment - guidance
Floors should be wood,
concrete or ceramic tile.
If floor are covered with
synthetic material, the
relative humidity should
be at least 30%.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
T
should be that of a
typical commercial or
hospital environment.