NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electro-
magnetic propagation is affected by absorption and reflection from struc-
tures, objects and people.
a. Field
strengths
tions
for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot
be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in
which the MASTER PALM 3® Handheld pulse oximeter is used exceeds the
applicable RF compliance level above, the MASTER PALM 3® Handheld
pulse oximeter should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such
as reorienting or relocating the MASTER PALM 3® Handheld pulse oximeter.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be
less than 3 V/m.
Recommended separation distances between portable and mobile RF
communications equipment and the EQUIPMENT or SYSTEM – for EQUIPMENT
or SYSTEM that are not LIFE-SUPPORTING
Recommended separation distances between
portable and mobile RF communications equipment and the MASTER
The MASTER PALM 3® Handheld pulse oximeter is intended for use in
an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the MASTER PALM 3® Handheld
pulse oximeter can help prevent electromagnetic interference by maintai-
ning a minimum distance between portable and mobile RF communications
equipment (transmitters) and the MASTER PALM 3® Handheld pulse oxime-
ter as recommended below, according to the maximum output power of the
communications equipment.
from
fixed
transmitters,
PALM 3® Handheld pulse oximeter.
111