IEC 60601
Immunity test
test level
40 % U
(60 % dip
in UT) for 5
cycles
70 % U
(30 % dip in
UT) for 25
cycles
< 5 % U
(> 95 % dip in
UT) for 5 sec
Power
frequency
(50/60Hz)
3 A/m
magnetic field
IEC61000-4-8
NOTE : U
is the a.c. mains voltage prior to application of the test level.
T
Guidance and manufacturer's declaration – electromagnetic immunity – for
EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration – electromagnetic immunity
The MASTER PALM 3® Handheld pulse oximeter is intended for use in the
electromagnetic environment specified below. The customer or the user of
MASTER PALM 3® Handheld pulse oximeter should assure that it is used in
such an environment.
Compliance
level
40 % U
T
T
(60 % dip
in UT) for 5
cycles
70 % U
T
T
(30 % dip in
UT) for 25
cycles
< 5 % U
T
T
(> 95 % dip in
UT) for 5 sec
3 A/m
109
www.spengler.fr
Electromagnetic envi-
ronment - guidance
Power frequency
magnetic fields Should
be at levels characteristic
of a typical location in
a typical commercial or
hospital environment.