Télécharger Imprimer la page

Spengler MASTER PALM 3 Manuel Utilisateur page 112

Oxymètre de pouls portable

Publicité

Separation distance according to frequency of transmitter
Rated
maximum
150KHz-80MHz
output
power of
d
transmitter
=
(W)
0,01
0,35
0,1
1,11
1
3,5
10
11,1
100
35
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in metres (m) can be estimated using
the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE 2: These guidelines
Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
80MHz-800MHz
3
5 .
p
d
=
 
 
V
1
0,12
0,37
1,17
3,69
11,67
may not apply in all situations.
112
www.spengler.fr
(m)
800MHz-2,5GHz
3
5 .
P
d
=
 
 
E
1
0,23
0,74
2,33
7,38
23,33
7
P
 
 
1
E

Publicité

loading