Rs 232 Interface Commands - Heidolph LABOROTA 20 S A Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

E

41. RS 232 interface commands

query
Command
Act commands
VAC?
VAP?
BATH?
CHILL?
SPEED?
TIME?
COUNT?
Safety functions
SC?
FS?
RS?
OP?
triggering
Command
Set commands
VAC=XXXX!
BATH=XXXX!
SPEED=XXX!
TIME=XXXX!
Actions
SVX!
VVX!
BVX!
SBX!
SSX!
LU!
LD!
action
Act Vacuum
Act Vapor
Act Bath
Act Chiller
Act Speed
Act Time
Act Time
Safety Cover
Flask Support
Rotation Stop
Over-
Temperature
Prot.
action
remarks
Set Vacuum
4-digit in mbar; range set 0000 to 1200
4-digit, in 1°C increments
Set Bath
range set 0000 to 1800 equals 0.0 to 180.0
Set Speed
3-digit, in rpm; range set 000 to 180
4-digit, in hh:mm
Set Timer
range set 0000 to 9959 equals 00:00 to 99:59
Start Stop Vacuum
SV0! = OFF ; SV1! = ON
Valve
vacuum valve; VV0! = CLOSE ; VV1! = OPEN
Ventilation
aerating valve; BV0! = CLOSE ; BV1! = OPEN
Start Stop Bath
SB0! = heater OFF ; SB1! = heater ON
Start Stop Speed
SS0! = rotation OFF ; SS1! = rotation ON
lift UP
lift DOWN
remarks
elapsed time, continuous operation
time, timer mode
0 = action; 1 = no action
0 = action; 1 = no action
0 = action; 1 = no action
0 = action; 1 = no action
112

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Laborota 20 s a oLaborota 20 e aLaborota 20 e a o

Table des Matières