AOIP TC-TM66 Serie Notice D'emploi page 70

Publicité

TC-TM66xx
P
ARAMETER SETTINGS
This function allows you to adjust the emission from the product's internal measurement gauge. This procedure takes several tens of seconds.
Measurement CSF:
This function allows you to adjust the Cold Junction Compensation (French abbreviation, CSF). You need to know precisely what the temperature of the reference
thermocouple used is, and the latter must be a type K thermocouple.
Example of a resistive sensor calibrator
Measurement:
Allows you to access the gauge adjustment functions.
st
The 1
screen indicates the gain and offset correction value found for the adjustment of this gauge. The counter indicates the number of adjustments the device
has undergone with the date of the last adjustment.
To perform an adjustment:
Press the VAL key.
The adjustment is performed in 2 points, around 010 Ohm and 300 Ohm. Apply the calibration resistance every time it is requested, and enter the value of this
calibration in the "Value Applied" field.
Confirm each step using the VAL key.
Offset ohms:
This function allows you to adjust the offset ohms in the measurement and simulation channel. Two steps are to be carried out insequence, following the
instructions
70/70