Using the navigation keys (↑ ↑ ↑ ↑ and ↑ ↑ ↑ ↑ ), choose the CSF (OFF: None, ON: internal or programmed).
•
Press VAL to confirm.
•
Press VAL (again) to confirm the desired function and go back to the measurement screen.
The Measurement mode makes it possible to display the Min (bottom left), Average (bottom centre) and Max values (bottom right) from the last RAZ
Min/Max command.
•
Access this command by pressing the F2 key.
•
Using the navigation keys (↑ ↑ ↑ ↑ and ↑ ↑ ↑ ↑ ), position the cursor in the RAZ Min/Max field.
•
Confirm the latter using the VAL key.
Note:
Following a significant thermal shock, it is advisable to allow the device to stabilise its temperature to use the internal reference junction (CSF)
with the utmost precision (TC6621, TM6602 and TM6630 units)..
C.1.2
Voltage/resistance or temperature simulation
To access the Emission mode:
•
The choice of Emission mode is made using the F2 key (mode menu).
•
Using the navigation keys (↑ ↑ ↑ ↑ and ↑ ↑ ↑ ↑ ), position the cursor in the Emission field going down the menu.
•
Confirm your choice using the VAL key.
Once you have confirmed the Emission mode, define the type of generation:
•
Continuous (manual or arrows or default editing).
•
Incremental (by step or "manual default").
•
By single ramp (only one ramp emitted).
•
By cyclic ramp.
•
Synthesiser ("automatic default").
Voltage/resistance or temperature generation/manual editing?
•
Press the F2 key to display the edit menu.
•
Using the navigation keys (↑ ↑ ↑ ↑ and ↓ ↓ ↓ ↓ ), choose the Manual edit mode and confirm (VAL key).
•
Press VAL again and enter your value using the navigation keys:
↑ ↑ ↑ ↑ and ↓ ↓ ↓ ↓ to increase or decrease the value
o
and
to select the digit to modify (hundreds/tens/unit/decimal place/hundreds).
o
Beware: if the Scaling mode is ON, the value to edit is expressed in the unit defined, otherwise this value is expressed in Volts or in °C or °F.