Programming parameters in Single Ramp mode
T0
Tdt
To : time pressing the generation start key
Td : Time frame
Tdt : Total Time
NB : Low Level
NH : High Level
The LOW level and HIGH level levels are expressed:
As a percentage of the gauge if the scaling mode is ON.
In Volt or in temperature units if the scaling mode is OFF and according to the type of value emitted (voltage or temperature emission).
The Total Duration corresponds to the amount of increase time required to go from the Low level to the High level (and vice versa in the case of a decrease).
It is expressed in seconds and the max time is limited to 1000s.
The Time frame corresponds to the amount of time you can have between pressing the emission start key and the actual starting of generation. It is expressed
in seconds and the max time is limited to 1000s.
Use the F2 key to access the CONFIGURATION/RAMP menu. Attention: the appropriate function mode must be programmed (INCREASES mode) to access
the CONFIGURATION/RAMP/SINGLE menu.
In the CONFIGURATION/RAMP/SINGLE menu, use the F2 key to go to the next field.
Use the navigation keys to enter the value:
As a percentage of the gauge if the scaling mode is ON.
↑ ↑ ↑ ↑ and ↓ ↓ ↓ ↓ to increase or decrease the value
    and
to choose the hundreds/tens/units/decimal place/hundreds/thousands.
Press VAL to save the parameters.
To quit the menu without saving, press CLEAR.
Cyclic ramp configuration?
The figure below illustrates the type of single ramp that can be generated and their parameters:
Programming parameters in Cyclic ramp mode
NH
NB
T0
Tb
Tm
Td
To : time pressing generation start key
Td : Time frame
Tb : Low duration
Tm : Rise (rise time)
Th : High duration
Tf : Drop (drop time)
NB : Low Level
NH : High Level
NH
NB
Td
T0
Tdt
Th
Tf
Td
T0
Tdt
D
65/65