AOIP TC-TM66 Serie Notice D'emploi page 64

Publicité

TC-TM66xx
R
ELATED
This is the number of cycles that need to be performed.
Time frame:
This is the time lapse between 2 repetitions.
D.5.2
Configuration of the ramp generation
The CONFIGURATION/RAMP menu is used for the generation of ramps by increases, single or cyclic.
Ramp by increases configuration?
The figure below illustrates the type of single ramp that can be generated and their parameters:
Programming parameters in Increases mode
INC
T0
Tdt
To : time pressing the generation start key
Td : Time frame
Tdt : Duration
NB : Low Level
NH : High Level
INC : Increase (value of increase as a %, in mVolt, °C or °F)
The LOW level and HIGH level levels are expressed:
As a percentage of the gauge if the scaling mode is ON.
In mVolt or in temperature units if the scaling mode is OFF and according to the type of value emitted (voltage or temperature emission).
The Duration corresponds to the amount of increase time required to go from the Low level to the High level (and vice versa in the case of a decrease). It is
expressed in seconds and the max time is limited to 1000s.
The Time frame corresponds to the amount of time you can have between pressing the emission start key and the actual starting of generation. It is expressed
in seconds and the max time is limited to 1000s.
Use the F2 key to move to the next field.
Use the navigation keys to enter the value:
As a percentage of the gauge if the scaling mode is ON.
↑ ↑ ↑ ↑ and ↓ ↓ ↓ ↓ to increase or decrease the value
    and
to choose the hundreds/tens/units/decimal place/hundreds/thousands.
Press VAL to save the parameters.
To quit the menu without saving, press CLEAR.
Single ramp configuration?
The figure below illustrates the type of single ramp that can be generated and their parameters:
64/64
F
UNCTIONS
NH
NB
Td
T0
Tdt
Td
T0
Tdt