Using the F1 function key, open the menu to access the following functions:
Init EEP:
Allows you to initialise part of the EEPROM (Calibrated Sensors Coefficient)
Example of a thermocouple calibrator:
Measurement:
Allows you to access the gauge adjustment function (stated as 100 mV).
st
The 1
screen indicates the gain and offset correction value found for the adjustment of this gauge. The counter indicates the number of adjustments the device
has undergone with the date of the last adjustment.
To perform an adjustment:
•
Press the VAL key.
The adjustment is performed in 2 points, around 0V and 100 mV. Apply the calibration signal every time it is requested, and enter the value of this calibration