TC-TM66xx
M
ODE
C.
M
P
ODE
ROGRAMMING
C.1.1
Voltage/resistance or temperature measurement
•
The choice of measurement or emission mode is made using the F2 key (mode menu).
•
Using the navigation keys (↑ ↑ ↑ ↑ and ↓ ↓ ↓ ↓ ), position the cursor in the Measurement field going down the menu.
•
Confirm your choice using the VAL key.
Note that the Measurement mode is the mode selected by default.
•
The function type selection (Thermocouple or resistive sensor type) is made using the F1 key (Configuration menu).
•
Using the navigation keys (↑ ↑ ↑ ↑ and ↓ ↓ ↓ ↓ ), position the cursor in the Function field.
•
Confirm the latter using the VAL key.
In the CONFIGURATION MESURE (MEASUREMENT CONFIGURATION) menu, position the cursor in the Couple or Sonde field using the F1 key.
•
Enter the Couple or Sonde menu using the F2 key.
•
Choose the type of thermcouple (K,T, J...), using the navigation keys (↑ ↑ ↑ ↑ and ↓ ↓ ↓ ↓ ) or the type of resistive sensor (PT50, PT100, PT200...).
•
Press VAL to confirm.
•
Using the F1 key, define the Unit by positioning the cursor on it.
•
Enter the menu by pressing F2.
•
Using the navigation keys (↑ ↑ ↑ ↑ and ↓ ↓ ↓ ↓ ), choose the unit.
•
Press VAL to confirm.
Attention, the choice of °C or °F is made in the Setup\Preferences\temp unit menu
•
Using the F1 key, define the CSF used by positioning the cursor on it (for TC6621, TM6602 and TM6630 units only).
•
Enter the menu by pressing F2.
52/52
P
ROGRAMMING