Instructions For Sterilizing The Outer Sheath And Tray Cradle, Cont.
MAINTENANCE
STORAGE
MANUFACTURER CONTACT
Instructions For Disinfecting All Other Parts
WARNINGS
LIMITATIONS ON
REPROCESSING
POINT OF USE
CONTAINMENT AND
TRANSPORTATION
CLEANING
DISINFECTION
DRYING
ENGLISH • 8
Visually inspect to ensure that all contamination has been removed.
Check for distortion, damage or wear. Discard damaged, worn or corroded instruments.
To maintain sterility, instruments should remain bagged until ready for use.
In the United States, contact DENTSPLY Professional Customer Service Technical Support
at 800-989-8826. For areas outside the United States, contact your local DENTSPLY
Division.
The charging base, inner module, foot pedal, power supply and double-ended
extension cable (not shown) are not sterilizable by autoclave, but can be disinfected
following the procedures listed below.
Only use water based non-immersion type disinfectant solutions.
Per the Centers for Disease Control and Prevention (CDC), chemical germicide
registered with the EPA as a "hospital disinfectant" and labeled for "tuberculocidal"
(i.e., mycobactericidal) activity is recommended for disinfecting surfaces that have
been soiled with patient material.
The following tuberculocidal disinfectant has been found to be effective in achieving a
desired level of disinfection
• Volo Wipes™ Surface Disinfectant (manufactured by Sultan Healthcare) or
equivalent
Repeated cleaning has minimum effect on these instruments. End of life is normally
determined by wear and damage due to use.
Do not use disinfectant solution on sterilizable outer sheaths. Refer to sterilization
procedures for sterilizable outer sheaths.
Remove excess soil with disposable cloth or paper wipe. Discard wipe after use.
Handle with care.
Generously spray disinfectant solution on a clean cloth. Wipe the outer surfaces of
the charging base, inner module, foot pedal, power supply and cords. Discard used
cloth. Wipe dry with a clean cloth.
Generously spray disinfectant solution on a clean cloth. Wipe the outer surfaces of the
charging base, inner module, foot pedal, and the power supply and its cord.
Discard used cloth.
Allow disinfectant to air dry.
When cleaning, wipe surfaces dry with a clean cloth. To achieve disinfection, allow
surfaces to air dry.
(Charging Base, Inner Module, Foot Pedal and Power Supply)