Télécharger Imprimer la page

Dentsply Sirona Midwest RDH Freedom Mode D'emploi page 10

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 20
Instructions For Disinfecting All Other Parts
MAINTENANCE
STORAGE
MANUFACTURER CONTACT
The instructions provided above have been validated by DENTSPLY as being capable of preparing a medical device for
re-use. It remains the responsibility of the processor to ensure that the processing is actually performed using equipment,
materials and personnel in the processing facility to achieve the desired result. This requires validation and routine
monitoring of the process. Likewise, any deviation by the processor from the instructions provided should be properly
evaluated for effectiveness and potential adverse consequences.
9. System Setup
Both the handpiece and foot pedal must be
charged prior to first use. Only use the Midwest
RDH Freedom power supply. Failure to do so
could cause your system to malfunction and
void your warranty. Charge the handpiece and
foot pedal for at least 90 minutes. (See Section 10,
Charging the Handpiece and Foot Pedal).
10. Charging the Handpiece and Foot
Pedal
When the corresponding indicator light is orange,
the handpiece and/or foot pedal must be charged.
Foot-pedal charge
indicator
10.1 A double-ended extension cable has been sup-
plied that allows you to charge both the inner
module and the foot pedal at the same time. Plug
the female onto the Power Supply Cable plug and
plug the male ends into the charging base and foot
pedal.
You can also charge the inner module and the foot
pedal individually using the power supply cable
alone.
Visually inspect to ensure that all contamination has been removed.
Visually inspect power supply and cords for damage.
Ambient temperature range: -20˚C to 50˚C
Relative humidity range: 45 - 95% (non-condensing)
In the United States, contact DENTSPLY Professional Customer Service Technical
Support at 800-989-8826. For areas outside the United States, contact your local
DENTSPLY Division.
Handpiece charge
indicator
(Charging Base, Inner Module, Foot Pedal and Power Supply),
Power supply
connector
10.2 Plug the power supply into a wall outlet.
NOTE: The green power-on indicator on the back of the
charging base illuminates when the charging base is
successfully connected.
NOTE: Unplug the power supply from the wall outlet
(power mains) to remove all power to the charging base.
PRECAUTION: Connect to single phase AC power 100-
240V power only. Otherwise, malfunction will occur.
Do not unplug the power supply by pulling on the cord.
Cont.
Power Indicator
ENGLISH • 9

Publicité

loading

Ce manuel est également adapté pour:

0060-gr dt