11.5 Invert the inner module and place it on the charg-
ing base as shown. The LED shown below will
blink green if the position is correct for synchroni-
zation.
11.6 When the indicator lights display a scrolling green
and orange pattern, press the red button in the
bottom compartment of the foot pedal within 15
seconds.
NOTE:
• The LED beside the synchronization switch of the foot
pedal flashes orange to indicate that the foot pedal is in
synchronization mode.
11.7 If the synchronization was successful, the three
indicator lights on the handpiece and the LED in
the foot pedal will flash green several times.
11.8 If all three handpiece indicator lights, and the foot
pedal LED flash orange, the syncing was unsuc-
cessful. Return the inner module to an upright
position in the charging base and repeat the proce-
dure.
12. Preparation for Use
12.1 Position the foot pedal for use, ensuring that the
floor is level.
LED
12.2 Ensure that the outer sheath has been sterilized
according to the Infection Control Procedures (sec-
tion 8).
12.3 When it is sufficiently charged, insert the inner
module into a Disposa-Shield
the nose of the inner module completely clears the
opening of the Disposa-Shield sleeve.
PRECAUTION: Do not install the covered inner module
into the outer sheath without ensuring the Disposa-
Shield sleeve is clear of the inner module nose. The
Disposa-Shield protects the inner module from debris
and splatter. Do NOT apply the Disposa-Shield to the
outer sheath, which must be sterilized before each use.
12.4 Install the inner module into the outer sheath by
aligning the swoops on each and snapping in
place.
12.5 Attach a NUPRO Freedom Disposable Prophy
Angle to the handpiece by aligning the swoops on
each and snapping in place.
12.6 Verify that all parts of the handpiece are securely
attached before use.
®
sleeve. Ensure that
Aligned correctly
Fully-Assembled Handpiece
ENGLISH • 11
Aligned correctly