Software operation
Set
7.5
Machine Menu
7.5.1
Querying system information
Parameter query
30 / 39
[STOP/OPEN] button.
2.
Press the
�� Precooling of the rotor is terminated.
Ramp-down takes place with the selected brake level.
The brake level is displayed.
The precooling speed is adjustable from 500 RPM up to the maximum rotor
speed in increments of 10 RPM. It is preset to 10000 RPM.
The rotor is stationary.
The lid is open.
1.
Press and hold the
't/min = X ' is displayed after 8 seconds.
��
[Cooling] button.
2.
Press the
�� Precooling speed
[Rotary knob] to set the precooling speed.
3.
Use the
[START/IMPULS] button.
4.
Press the
�� The setting is stored.
'*** ok ***' is displayed briefly.
[STOP/OPEN] button twice or wait 8 seconds to exit the
5.
Press the
menu.
The rotor is stationary.
1.
Press and hold the
'SOUND/BELL' is displayed.
��
[SELECT] button repeatedly until 'FU/CCI – S.' is displayed.
2.
Press the
Program version for the frequency converter
3.
Press the
[SELECT] button repeatedly until 'HOURS' is displayed.
Internal operating hours (the time during which the centrifuge was
switched on)
4.
Turn to the right with the
'STARTS' is displayed.
��
Number of centrifugation runs
5.
Turn to the right with the
'ROTORCHG1' is displayed.
��
Internal operating hour of the last rotor change
6.
Turn to the right with the
'ROTORCHG2' is displayed.
��
Internal operating hour of the penultimate rotor change
7.
Turn to the right with the
'OPhoursCHG' is displayed.
��
Internal operating hour of the last operating hours change
8.
Turn to the right with the
'IMBALCHG' is displayed.
��
Internal operating hour of the last imbalance cut-off change
Rev.: 15 / 01.2024
[Cooling] button.
'RPM = XXXX' is displayed.
[SELECT] button for 8 seconds.
[Rotary knob] .
[Rotary knob] .
[Rotary knob] .
[Rotary knob] .
[Rotary knob] .
AB2200en