POWER SUPPLY and CONNECTORS
WIRING and POWER SUPPLY RANGE
MICRA-M
85 V – 265 V AC 50/ 60 Hz
MICRA-M6
22 – 53 V AC 50/ 60 Hz
Borne 1: Phase
Borne 2: Neutral
NOTE: When DC power supply (direct),
polarity in connector CN1 is indistinct.
Instrument frontal view
KOSMOS SERIE
WARNING: If not installed and used in accordance with these instruc-
tions, protection against hazards may be impaired.
In order to guarantee the electromagnetic compatibility, the following guide-
lines should be kept in mind:
Power supply wires may be routed separated from signal wires.
Never run power and signal wires in the same conduit.
Use shielded cable for signal wiring and connect the shield to the ground of the
indicator. The cables section should be ≥0.25 mm
INSTALLATION
To meet the requirements of the directive EN61010-1, where the unit is perma-
nently connected to the mains supply, it is obligatory to install a circuit breaking
device easy reachable to the operator and clearly marked as the disconnect
device.
or 100 – 300 V DC
or 10,5 - 70 V DC
www.ditel.es
CONNECTORS
CN1 To perform wiring connections, strip the wire leaving
from 7 and 10 mm exposed to air and insert it in the proper
terminal while pushing the fingertip down to open the clip
inside the connector as indicated in the figures.
Each terminal accepts cables of section between 0.08 mm²
and 2.5 mm² (AWG 26 ÷ 14).
CN2 & CN3 To perform wiring connections, strip the wire
leaving from 5 and 6 mm exposed to air and insert it in the
proper terminal while pushing the fingertip down to open
the clip inside the connector as indicated in the figures.
Each terminal accepts cables of section between 0.08 mm²
and 0.5 mm² (AWG 28 ÷ 20).
MICRA-M
2
72