Error Messages On The Display - Bühler technologies TC-Double Notice De Montage

Masquer les pouces Voir aussi pour TC-Double:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 18
TC-Double (+)
6 Maintenance
Always observe the applicable safety- and operating regula-
tions when performing any type of maintenance. Please refer
to the original operator's manual on the included CD or online
at www.buehler-technologies.com for maintenance informa-
tion.
7 Service and Repair
Please refer to the original operator's manual on the included
CD or online at www.buehler-technologies.com for a detailed
description of the unit including information on
troubleshooting and repair.

7.1 Error messages on the display

If an error occurs, the display will read "Err". Press the "
button to show the error number(s).
Error messages will appear until the unit has been restarted
or the error is cleared using the "Func" button. It can only be
cleared if the cause for the error has been corrected.
Causes / Action: The following is a list of the most common
causes and actions for the respective error. If the actions listed
do not resolve the problem, please contact Service.
Problem /
Possible cause
Malfunction
No
• No voltage
display
• Loose connect-
ing cable
• Display defect-
ive
D1.02
(The software ver-
sion for the display
(perman-
will appear).
ent)
• Not commu-
nicating with
the controller
Error
• An error has
occurred
Error 01
• Controller
malfunction
• Malfunction
controller 2
Error 03
• Microcontrol-
ler Fault /
MCP2
Error 04
• EEPROM error
• Fan malfunc-
tion / fan out
of order
16
Bühler Technologies GmbH
"
Action
• Check the supply
cable
• Check fuse
• Check connec-
tions
• Check connec-
tions
• Read the error
number as de-
scribed above
• Clear error (tem-
porary fault)
• Disconnect from
power for approx.
5 s
• Contact service
• Clear error (tem-
porary fault)
• Disconnect from
power for approx.
5 s
• Contact service
• Contact service
• Contact service
• Check fan
(blocked?)
• Check connection
Error 22
• Moisture de-
tector 1 cable
break
Error 32
• Moisture de-
tector 2 cable
break
Error 40
• General error
temperature
sensor 1
Error 41
• Low temperat-
ure / short-cir-
cuit temperat-
ure sensor 1
Error 42
• Excess tem-
perature /
short-circuit
temperature
sensor 1
Error 43
• Measurement
fluctuation
temperature
sensor 1
Error 50
• General error
temperature
sensor 2
Error 51
• Low temperat-
ure / short-cir-
cuit temperat-
ure sensor 2
Error 52
• Excess tem-
perature /
short-circuit
temperature
sensor 2
Error 53
• Measurement
fluctuation
temperature
sensor 2
• Check moisture
detector line
• Check moisture
detector
• Check moisture
detector line
• Check moisture
detector
• Sensor possibly
defective
• Check temperat-
ure sensor con-
nection
• Check temperat-
ure sensor con-
nection
• Check temperat-
ure sensor con-
nection
• Sensor possibly
defective
• Check temperat-
ure sensor con-
nection
• Check temperat-
ure sensor con-
nection
• Check temperat-
ure sensor con-
nection
BX440027 ◦ 10/2018