If an alarm message is displayed the cycle is to be considered aborted: it will
be necessary to repeat all the preparation and sterilization operations.
If an error message appeared on the display during the sterilization phase, the
cycle failed and the instruments in the chamber were not sterilized.
MESSAGE
ERROR100
ERROR120
ERROR125
ERROR140
ERROR199
ERROR400
ERROR401
ERROR450
ERROR700
ERROR760
ERROR800
If the problem is coming out even if you carried out the listed operation, please contact the
MESSAGES OF ERRORS
To RESET the error message, push the button 11
The sterilization cycle was interrupted by the operator pressing the
button 10.
Low pressure in the chamber while sterilizing
High pressure in the chamber while sterilizing
The pressure did not increase properly during the preheating phase or
the machine could not discharge the pressure at the end of the
sterilizing cycle. If this is happening even if the water is in the chamber,
dry and empty the sterilization chamber, reset the alarm, let the water in
and restart the cycle up
The power was disconnected while the sterilization cycle
The cycle was started up without let water in the chamber.
Dry and empty the sterilization chamber, reset the alarm, let the water in
and restart the cycle up.
The cycle was started up but 5 min went from the water inlet in the
chamber. Dry and empty the sterilization chamber, reset the alarm, let
the water in and restart the cycle up
The cycle was started up but the machine was hot. After the door
closing the machine let the water out of the chamber.
Reset the alarm, wait until when the machine is cold, Dry and empty the
sterilization chamber, let the water in and restart the cycle up
During the preheating phase, in 40 minutes was not reached pressure
value established for sterilization. If this is happening even if the water is
in the chamber, dry and empty the sterilization chamber, reset the alarm,
let the water in and restart the cycle up
Failure in the device for the pressure level
The pressure did not go down properly during the drying phase.
CAUSE
assistance