Télécharger Imprimer la page

Tecno-gaz STERIL BEAUTY Mode D'emploi page 32

Publicité

Les langues disponibles

Les langues disponibles

On the USB memory is saved a text file with the follow formatting:
CYCLE NUMBER
CYCLE TYPE
CYCLE ISSUE
ANY ERROUR CODE
TECHNICAL PARAMETERS
10
Correct maintenance of the autoclave assures correct funcioning of it and a secure saving
in terms of time and costs for assistance and maintenance. The following operations are
compulsory feasible by operators.
Cleaning of chamber
Clean periodically the chamber, remove eventual deposits or debris, thus avoiding the
introduction in the discharging circuit of material which can cause obstructions. For
correct cleaning use only deminerlized water and the abrasive sponge provided.
To execute absolutely with cold chamber to avoid burns – Never use solvents,
detergents, chemical solutions, descaling agents or other similar products.
Cleaning of chamber filter
Pull up the filter, paying attention not to damage it, wash with demineralized water and
dry with cloth. Then replace in the seat, making sure that protrudes about 15 mm.
Cleaning of tray and tray holder
Clean with sponge supplied (non abrasive side) and demineralized water.
Cleaning of the door gasket
Periodically remove eventual residuals which are deposited on the circumference of the
seal using water and the sponge provided (non abrasive side), or else a moistened cloth.
MAINTENANCE
Every 20 cycles or one a week
Every 20 cycles or once a week
Every 20 cycles or once a week
Every 20 cycles or once a week
ENGLISH
13

Publicité

loading

Ce manuel est également adapté pour:

As004zsm