Turn on the tap 08 acting to the ring nut turning in couterclockwise.
-
Insert the pipe with funnel and spill the demineralized water (watch water features in
the following table).
-
Spill all the water in to the funnel.
IMPORTANT: Turn on the tap 08 acting to the ring nut turning in clockwise after the load,
in order to avoid steam releases to the frontal part.
Evaporation residue
Silicon oxide
Iron
Cadmium
Lead
Remains of heavy metals apart from iron, cadmium, lead
Chloride (Cl')
Phosphate (P20s)
Conductivity (at 20°C)
Ph value (acidity level)
Appearance
Hardness (and alkaline earth ions)
8
WATER FEATURES TO USE
10
£
1
£
0.2
£
0.005
£
0.05
£
0.1
£
2
£
0.5
£
15
£
5 ¸ 7,5
Transparent, clear, without deposits
0.02
£
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
ms/cm
mmol/l